  • 1q(1775724341991).webp

NDX6000 Series Fully Automatic Anti-Interference Dielectric Loss Tester



Contact phone number:

Product Details


■ Product Overview

Dielectric loss measurement is a fundamental method in insulation testing, capable of effectively detecting overall moisture ingress and degradation of electrical equipment insulation. , as well as localized defects. It is widely used in electrical manufacturing, electrical equipment installation and handover, and preventive testing. Measuring the dielectric loss of transformers, instrument transformers, reactors, capacitors, bushings, surge arresters, and other components is the most fundamental method for assessing their insulation performance. By employing variable-frequency power‑supply technology and leveraging microcontrollers and modern electronic techniques, automatic frequency conversion and analog… / Number conversion and data operations , It achieves strong anti-interference capability, fast testing speed, high accuracy, fully automated digitalization, and simple operation. ; The power supply uses a high-power switching power supply. , Output 45Hz and 55Hz Pure sine wave , Automatic pressurization , Can provide up to 10kV the voltage ; Automatic filtering 50Hz Interference—suitable for on-site testing in environments with significant electromagnetic interference, such as substations.

■ Technical Specifications

Accuracy

Cx: ± ( Reading x1%+1pF)

tan delta : ± ( Reading x1%+0.00040)

Anti-interference performance indicator

Variable-frequency anti-interference, in 200% It can still achieve the aforementioned accuracy under interference.

Capacitance range

Internal high pressure

3pF-60000pF/10kV 60pF~1 mu F/0.5kV

External high voltage application

3pF~1.5 mu F/10kV60pF-30uF/0.5kV

Resolution

0.001pF4 significant figures

lg5 Scope

Unlimited, resolution 0.001% , Automatic identification of three types of test samples: capacitors, inductors, and resistors.

Test current range

10 mu A~1A

Internal high pressure

Set the voltage range

0.5~10kV

Maximum output current

200mA

Buck-boost method

Continuous smooth adjustment

Frequency accuracy

± 0.01HZ

Test frequency

4550556065Hz Single frequency

45/55H255/65Hz47.5/52.5Hz Automatic dual-frequency inverter

External high voltage application

Maximum test current in positive wiring 1A , power frequency or variable frequency 40-70Hz

Maximum test current during reverse wiring 10KV/1A , power frequency or variable frequency 40-70Hz

CVT Self-excitation method low-voltage output

Output voltage 3~50V , output current 3~30A

CVT Turns Ratio Measurement

Turns-ratio measurement accuracy

± reading x1%

Turns-ratio measurement range

10~99999

Phase measurement

Phase measurement accuracy

± 0.1 °

Phase measurement range

0-359. 9 °

Measurement time

Approximately 40s , related to the measurement method

Computer interface

Standard RS232 Interface

Input power supply

180V~270VAC 50Hz ± 1% , powered by mains electricity or a generator


Service system


Offering inspection and reception services

We guarantee to welcome your visit at any time and provide you with every possible convenience.

Provide a reasonable quotation.

We guarantee to provide a unified, reasonable quotation for the products you request within 24 hours.

Provide professional technical support

We guarantee that a technical specialist will respond to any technical questions you may have within 24 hours.

Provide detailed information

We guarantee that, within 24 hours, the product catalogues, software CDs, and other technical materials you require will be delivered to your company by our local branch or office.

Related products


Power Technology


Leave a message for inquiry


Welcome to leave us an online message; we will get back to you promptly.

%{tishi_zhanwei}%