ND3020 Series Multifunctional Electrical Meter Calibration Bench
Contact phone number:
Product Center
-
Series Resonant Withstand Voltage Test Device
-
High-Voltage Withstand Voltage Test Equipment Series
-
Transformer Testing Series
-
Insulating Oil and SF6 Gas Testing Equipment
-
Insulation Resistance and Grounding Resistance Testing Series
-
Instrument transformer, relay, and pressure relief valve testing series
-
Switchgear, Circuit Breaker, and Surge Arrester Testing Series
-
Battery and DC System Testing Series
-
Generator and Motor Testing Series
-
Cable and Overhead Line Inspection Series
-
Secondary Protection Testing Series
-
Partial Discharge-Free Detection and Live On-Line Monitoring Series
-
Power Quality and Safety Tools Testing Series
-
Other
Product Details
II. Main Technical Specifications and Performance:
1. Instrument class: Class 0.05
2. Computer communication port: RS-232 serial port
3. Technical parameters of the three-phase source:
3.1 Output Voltage:
3 * 57.7V Y 3*100V ∆
3 * 220V Y 3*380V ∆
Adjustable amplitude, continuously variable from 0% to 120%
Fineness adjustment: better than 0.01%
3.2 Output Current: 25 mA, 0.1 A, 1 A, 5 A, 25 A, 100 A
Adjustment range: continuously adjustable from 0% to 120%; adjustment resolution better than 0.01%.
Minimum output current: 5 mA (accuracy: <5%)
3.3 Output Power: (Voltage Circuit) 3 × 100 VA
(Current circuit) 3 × 150 VA
3.4 Output voltage, current, and power stability;
≤0.05%/120S
3.5 Output voltage and current waveform distortion: ≤ 0.5%
3.6. Three-phase symmetry: better than 120° o Soil 0.2 o
3.7 Frequency range: 45 Hz to 65 Hz; Adjustment resolution: 0.01 Hz
3.8 Phase shift range: -180.0 o ~180.0 o Fineness adjustment: 0.01 o
3.9 Indicating instrument class: The accuracy of voltmeters and ammeters is ±0.01%.
The phase meter has an accuracy of ±0.01. o
The frequency meter has an accuracy of ±0.01 Hz.
4. Technical parameters of the built-in three-phase standard meter:
4.1 Voltage Input
Input range (V): 40–480
Input range (V): 60, 120, 240, 480
Overload Capacity (V): 600
Input load ≤ 0.5 mA (480 V)
4.2 Current Input
Input range (A): 0.01–100
Input range (A): 100, 50, 25, 10, 5, 2.5, 1, 0.5, 0.25, 0.1, 0.05, 0.025, 0.01, 0.005
Overload Capacity (A): 150
4.3 Accuracy
4.3.1 Accuracy Class: 0.05 class (0.025 A to 100 A)
4.3.2 Power Measurement Accuracy
Active power measurement accuracy: 0.05% (PF ≥ 0.5)
Reactive power measurement accuracy: 0.1% (PF ≥ 0.5)
4.3.3 Accuracy of Electrical Energy Measurement
Active energy measurement accuracy: 0.05% (PF ≥ 0.5)
Reactive energy measurement accuracy: 0.1% (PF ≥ 0.5)
5. The device features comprehensive protection functions: voltage short-circuit protection, current open-circuit protection, and overload protection for the load.
Monitor overvoltage protection.
6. Supports harmonic tests from the 2nd to the 21st order, as well as voltage interruption and sag tests; harmonic amplitude is adjustable.
7. Power supply: 230 V ±10% ~ 50 Hz
8. Maximum power consumption: 300 VA
9. Operating environment: Temperature 20°C ± 5°C, relative humidity ≤ 85%
Service system
Offering inspection and reception services
We guarantee to welcome your visit at any time and provide you with every possible convenience.
Provide a reasonable quotation.
We guarantee to provide a unified, reasonable quotation for the products you request within 24 hours.
Provide professional technical support
We guarantee that a technical specialist will respond to any technical questions you may have within 24 hours.
Provide detailed information
We guarantee that, within 24 hours, the product catalogues, software CDs, and other technical materials you require will be delivered to your company by our local branch or office.
Related products
Power Technology
2016-04-05
Leave a message for inquiry
Contact Us
National Service Hotline: 400-027-8938
Mobile (Manager Chen): 133 9714 9829
Address: Building No. 5, Donghu High-tech Industrial Innovation Base, No. 9 Miaoshan Avenue, Miaoshan Development Zone, Jiangxia District, Wuhan City
Scan it
Follow the official mobile website