Product Center
MengziND2678 Series Water-Internal-Cooling Generator Insulation Resistance Tester
Category:
MengziGenerator and Motor Testing Series
Phone: 0512-66101666
Product Description
■Product Overview
The water‑cooled generator insulation resistance tester is specifically designed for insulation measurements on water‑cooled generators, and it can also be used in laboratories or field settings to perform insulation tests on other materials.
The instrument’s output current is greater than 20mA , maximum output voltage 5000V It features a high-precision microcurrent measurement system and a digital step-up voltage system. Measurement can be performed by connecting only one high-voltage lead and one signal lead to the device under test.
Measurements are performed automatically, with results displayed on a large LCD screen and stored for later retrieval.
Product Features:
● Automatically compensates and adjusts the water‑induced polarization potential. The resistance between the base and the water‑collecting pipe is as low as 3k Ω( 2500V ) /6k Ω( 5000V ), which can also ensure measurement accuracy.
● High output power; the load resistance between the circuit and the manifold can be as low as 100k Omega (2500V For 75k Omega ), Insulation resistance measurement can be as low as 1M Omega.
■ Key Features
. Key features
1 , adopt 32 Controlled by a microcontroller, with a fully Chinese-language user interface for convenient operation.
2 Automatically calculates the absorption ratio and polarization index, and stores the results automatically. 15 seconds, 1 minutes, 10 The insulation resistance value at a given minute, facilitating analysis.
3 , large output current, (2500V 、5000V The output below is greater than 20mA) 。
4 The high-voltage generation module employs a fully enclosed design and incorporates internal protective resistors, ensuring safety and reliability.
5 It boasts strong anti-interference capabilities and can meet the operational requirements of ultra-high-voltage substations on-site.
6 , Automatically discharges upon completion of testing and monitors the discharge process in real time.
■ Product Specifications
Rated voltage | 2500V、5000V |
Measurement range | Insulation resistance 1M Ω~ 40G Omega Absorption ratio, polarization index 1 ~ 19.99 |
Margin of error | ± 10%RDG , (when 1M Ω~ 5G Ω) ± 20%RDG , (when 0.1M Ω~ 1M Ω; 5G Ω~ 40G Ω) |
Operating voltage | Rated voltage × ( 1 ± 10% ) |
Short-circuit current | > 20mA |
Insulation resistance | 50M Ω( 1000V ) (between the measurement circuit and the enclosure) |
Durability Pressure | AC 5kV 50Hz 1min (Between the measurement circuit and the enclosure) |
Power supply | 180 ~ 270VAC,50Hz/60Hz ± 1% |
Work environment | Temperature 0 ~ 40 °C, relative humidity 20 ~ 80% |
External dimensions | 320 × 240 × 135mm3 |
Heavy Quantity | 5kg |
Contact Us
National Service Hotline: 400-027-8938
Mobile (Manager Chen): 133 9714 9829
Address: Building No. 5, Donghu High-tech Industrial Innovation Base, No. 9 Miaoshan Avenue, Miaoshan Development Zone, Jiangxia District, Wuhan City
Scan it
Follow the official mobile website